Snare technique

Coronary Rotablator complications

There are several techniques typically employed to retrieve a stuck rotablator burr. The snare can be be attempted but not in first attention.

This technique is feasible with a GooseNeck snare. Feed loop of the Goose Neck snare over the shaft of the rota burr and pull.

Goose neck snare angle

Return to main page

Disclaimer

This case report does not reflect the opinion of PCR or PCRonline, nor does it engage their responsibility.